Calibration: Avery Berkel L126
If your scale is showing a "Zero Capture" error (often indicating the cell is out of its 2% tolerance), you may need to perform a simple reset without a full span calibration: Access the micro-switch on the circuit board.
You will typically need weights equal to one-third, two-thirds, or full capacity (often 5kg, 10kg, or 15kg depending on your specific model).
Press the Zero button twice until the display shows "CAL" (Calibration). Confirm your selection by pressing the Test or Store button. avery berkel l126 calibration
The Avery Berkel L126 is a robust retail scale and indicator system used widely for trade-legal weighing. To maintain its accuracy and compliance with trade standards, regular calibration is essential. This guide covers the procedures for calibrating and resetting the L126, based on technical service standards. Before starting, ensure you have the following ready:
Use the Zero button to toggle between units (e.g., kilograms or ounces) and select your preference with the Test button. If your scale is showing a "Zero Capture"
Ensure the scale platform is empty and clean. The display will prompt "Load Zero" . Press the Test button to capture the zero point. The scale will scroll for a few seconds as it stabilizes.
Turn the scale on and allow it to sit for at least 15 minutes to stabilize the electronics. Step-by-Step Calibration Procedure Confirm your selection by pressing the Test or Store button
When the display flashes , press the Zero key until it flashes "E" .
The display will prompt for a weight (e.g., "Load 10" ). Gently place your certified 10lb (or equivalent metric) weight onto the center of the platform. Press the Test button again.